18
−
Illustration
Procedure
When the optical axis is shifted by 1mm or
more in the right-and-left direction:
Loosen the 3 screws CR2×4S and move the one
side of the mirror holder (1011) up and down
for adjustment.
After adjusting accuracy, tighten the 3 screws
CR2×4S securely and apply the screw lock.
When the optical axis is shifted by 1mm or
more in the up-and-down direction:
Loosen the 1 piece 1N2.6, adjust 1 piece
U2.6×6S and bring the optical axis to the cen-
ter.
CR2×4S:3 pcs.
Mirror holder
(1011)
5
1N2.6
Scale line
U2.6×6S:1 pc.
φ
0.2
aperture
Adjust the mirror holder in the up-and-down direction
and bring it to the center in the right-and-left direction.
7
5
6
7
P-SLD_OPTION-0-0308-14