5.1 

Headrest Locking Tension

 

If  the  position  of  the  chair  headrest  has  begun  to  move  with  the  clamp-locking 
handle  in  the  locked  position,  the  headrest  clamp  may  require  an  adjustment. 
Follow the instructions below to adjust the headrest assembly. 
 

Headrest Clamp Adjustment

 

5.1.1  Release the clamp-locking handle. Loosen the Phillips screw in the side of 

the clamp body, and then turn the slotted adjustment screw in the end of 
the clamp assembly to adjust the  clamping tension. (Clockwise for more 
tension, counter-clockwise for less tension). Adjust the clamp tension so 
that the clamp-locking handle closes easily and the pivots at each end of 
the clamp are secure. Repeat the adjustments until the proper adjustment is 
achieved.  Re-tighten  the  Phillips  screw  in  the  side  of  the  clamp  body  to 
secure the adjustment. 

  

Pillow Mount Pivot Adjustment

 

5.1.2  Turn the small adjustment screw in the pillow mount clamp to adjust the 

pillow  mount  clamp  tension  (Clockwise  for  more  tension,  counter-
clockwise  for  less  tension.  The  clamp  is  properly  adjusted  when  the 
headrest  pillow  position  can  be  adjusted  easily,  but  has  enough  bind  to 
remain in the position that it is set to after adjustment. 

 

5.2 

Swivel Brake Adjustment

 

If  the  chair  can  be  rotated  while  the  swivel  brake  handle  is  in  the  activated 
position (parallel to the floor) an adjustment may be required. The swivel brake 
adjustment screw is located behind the lift assembly on the underside of the metal 
chair  seat.  If  the  swivel  brake  clamping  tension  is  too  loose,  move  the  swivel 
brake  handle  to  the  released  position  (perpendicular  to  the  floor)  and  turn  the 
adjustment  screw  with  an  Allen  wrench  (Clockwise  for  more  tension,  counter-
clockwise  for  less  tension)  until  the  proper  adjustment  is  achieved.  The  swivel 
brake  is  properly  adjusted  when  the  chair  rotates  freely  with  the  swivel  brake 
released but does not rotate when the brake is activated. 
 

5.3 

Armrest Adjustment

 

The  armrests  may  require  periodic  adjustment  to  maintain  proper  function. 
Remove  the  three  arm  upholstery  mounting  screws,  and  then  remove  the  arm 
upholstery  to  gain  access  to  the  arm  adjustment  screw.  Turn  the  flat  head 
adjustment screw in the back side of the armrest casting with an Allen wrench to 
adjust the amount of play in the armrest latching mechanism (clockwise for more 
play,  counter-clockwise  for  less  play).  The  armrest  latching  mechanism  is 
adjusted  properly  when  the  mechanism  engages  properly  when  the  armrest  is 
down, but has minimal vertical play once engaged. Reinstall the arm upholstery 
after performing the adjustment. 
 
The  armrest  height  and  relative  position  can  be  adjusted.  Remove  the  backrest 
upholstery to gain access to the adjustment screws. Tighten/loosen the opposing 

Содержание OC-2300

Страница 1: ...i ...

Страница 2: ...fe condition if a malfunction occurs After Operation Never disconnect any plugs by pulling on the cable Doing so may cause damage to the internal connections If the chair has been idle for a period of time check all functions and safety features for normal operation prior to reuse Safety Precautions The OC 2300 exam chair conforms to UL 2601 1 and CAN CSA C22 2 No 601 1 M90 There is no disposal of...

Страница 3: ...4 4 0 OPERATION 4 4 1 Elevation Control 4 4 2 Motorized Recline Control 4 4 3 Auto Return Control 4 4 4 Memory Controls 5 4 5 Cancel Button 5 4 6 Swivel Brake Operation 5 4 7 Armrest Retraction 5 4 8 Headrest Position 5 4 9 Footrest Retraction 5 5 0 ADJUSTMENTS 6 5 1 Headrest Locking Tension 6 5 2 Swivel Brake Adjustment 6 5 3 Armrest Adjustment 6 6 0 MAINTENANCE 7 6 1 Paint and Nonmetallic Parts ...

Страница 4: ...Height retracted 22 5 571mm 20 0 508mm Seat Height full stroke 35 0 889mm 29 0 737mm Footrest Height 3 5 89mm 1 5 38mm Full Recline Length 75 1905mm Mode of Operation Continuous Use Type of Protection Class 1 protection Against Electrical Shock Ingress of Water Protection Ordinary Protection IPX0 AP or APG Not AP or APG equipment Electrical Voltage Current Frequency US Versions 120VAC 10 7 A 60 Hz...

Страница 5: ...Degrees C Relative Humidity Range 30 to 95 Atmospheric Pressure Range 10 1 to 15 4 psi 700 to 1060 hPa Acceptable Storage Time in operating environment 5 years in extreme environment 20 days Denotes dimensions for the High Low lift configurations SYMBOLS FOUND ON THE OC 2300 CHAIR Protective Earth ground location CAUTION Electrical shock possibility ATTENTION Information provided in manual Alterna...

Страница 6: ... appropriate wall outlet or into the ophthalmic stand 2 Connect the chair control cable to the ophthalmic stand and check the vertical movement of the ophthalmic chair 3 4 Alternate Connections When the Ophthalmic Chair model OC 2300 is not used in conjunction with a Topcon ophthalmic stand use the table below to configure the electrical inputs correctly Stand Cable Wire Code Pin Number Wire Funct...

Страница 7: ...nt of the full power recline ophthalmic chair model OC 2300 is made by pressing the appropriate arrow button on either membrane switch located on the sides of the back or with an optional footswitch The chair may also be controlled by the UP DOWN and AUTORETURN buttons on the exam stand 4 2 Motorized Recline Control The motorized recline buttons located on the sides of the backrest recline and upr...

Страница 8: ...ditional safety feature pressing any button on either chair membrane switch will also terminate any automatic motion 4 6 Swivel Brake Operation To rotate the chair move the swivel brake handle to the released position perpendicular to the floor rotate the chair to the desired position then move the swivel brake handle to the activated position parallel to the floor to secure the chair in position ...

Страница 9: ...e swivel brake handle is in the activated position parallel to the floor an adjustment may be required The swivel brake adjustment screw is located behind the lift assembly on the underside of the metal chair seat If the swivel brake clamping tension is too loose move the swivel brake handle to the released position perpendicular to the floor and turn the adjustment screw with an Allen wrench Cloc...

Страница 10: ...e chair from the power source WARNING Any service and or repairs other than replacement of the two exter nal fuses shall be performed only by a qualified technician The main power fuses are located in the rear of the base assembly in the LH side plastic shroud Unplug the chair from the AC power source Twisting the fuse cap counter clockwise and simultaneously pressing down easily replace the fuses...

Страница 11: ...8 If problems persist call your dealer or Topcon for assistance Topcon Medical Systems Inc 37 West Century Road Paramus NJ 07652 800 223 1130 www topconmedical com ...

Отзывы: