1) 

If  the  chair  has  a  1

st

  generation  circuit  board,  remove  the  blue  jumper  pin 

located between the transformer and the 9-pin connector. Re-install the base 
shroud. 

OR 

2) 

If the chair has a 2

nd

 generation circuit board, move the red jumper pin located 

between  the  9-pin  connector  and  the  6-pin  connector  labeled  ‘Stand  Safety’ 
from the OFF position to the ON position. Re-install the base shroud. 

 

3.5 

Power Up

 

 

 
Plug  the  chair  power  cord  into  a  wall  receptacle  or  into  the  exam  stand  or  cabinet 
system. The chair control membrane  switches should illuminate to indicate that the 
chair  is  operational.  Perform  a  test  of  all  chair  features  to  verify  proper  function. 
Refer to the operation section of this document for details. 

 
 

4.0 

SYSTEM OPERATION

  

The OC-2300's sturdy design and motorized recline features make it an excellent choice for 
Ophthalmic and Optometric exam rooms. 
 
Each of the chair control membrane switch buttons performs a specific function. Refer to 
Performance and Specifications for the button symbols and their corresponding functions. 

 

4.1 

Elevation Control 

Height  adjustment  of  the  full  power  recline  ophthalmic  chair,  model  OC-2300,  is 
made by pressing the appropriate arrow button on either membrane switch located on 
the sides of the back or with an optional footswitch. The chair may also be controlled 
by the UP, DOWN and AUTORETURN buttons on the exam stand. 

 

4.2 

Motorized Recline Control 

The  motorized  recline  buttons,  located  on  the  sides  of  the  backrest,  recline  and 
upright the backrest. To recline the patient, press the back arrow button on one of the 
membrane  switches.  Release  the  button  once  the  desired  position  is  achieved.  You 
may also preset one or both of the memory buttons to a predetermined condition. To 
return the patient to the upright position press the forward arrow button on one of the 
membrane switches. Release the button once the chair is upright. You may also press 
the  AUTORETURN  button  to  bring  the  chair  fully  upright  and  fully  down.  An 
optional dual function footswitch may also be used to recline the chair. 
 

4.3 

Auto-Return Control

 

An AUTORETURN button is located on each membrane switch and on most exam 
stands. Pressing the AUTORETURN button will automatically return the chair to the 
fully down and fully upright position. 

 
 

WARNING: Check all power cords and electrical connections for damage prior to

powering up the system.

!

Содержание OC-2300

Страница 1: ...i ...

Страница 2: ...fe condition if a malfunction occurs After Operation Never disconnect any plugs by pulling on the cable Doing so may cause damage to the internal connections If the chair has been idle for a period of time check all functions and safety features for normal operation prior to reuse Safety Precautions The OC 2300 exam chair conforms to UL 2601 1 and CAN CSA C22 2 No 601 1 M90 There is no disposal of...

Страница 3: ...4 4 0 OPERATION 4 4 1 Elevation Control 4 4 2 Motorized Recline Control 4 4 3 Auto Return Control 4 4 4 Memory Controls 5 4 5 Cancel Button 5 4 6 Swivel Brake Operation 5 4 7 Armrest Retraction 5 4 8 Headrest Position 5 4 9 Footrest Retraction 5 5 0 ADJUSTMENTS 6 5 1 Headrest Locking Tension 6 5 2 Swivel Brake Adjustment 6 5 3 Armrest Adjustment 6 6 0 MAINTENANCE 7 6 1 Paint and Nonmetallic Parts ...

Страница 4: ...Height retracted 22 5 571mm 20 0 508mm Seat Height full stroke 35 0 889mm 29 0 737mm Footrest Height 3 5 89mm 1 5 38mm Full Recline Length 75 1905mm Mode of Operation Continuous Use Type of Protection Class 1 protection Against Electrical Shock Ingress of Water Protection Ordinary Protection IPX0 AP or APG Not AP or APG equipment Electrical Voltage Current Frequency US Versions 120VAC 10 7 A 60 Hz...

Страница 5: ...Degrees C Relative Humidity Range 30 to 95 Atmospheric Pressure Range 10 1 to 15 4 psi 700 to 1060 hPa Acceptable Storage Time in operating environment 5 years in extreme environment 20 days Denotes dimensions for the High Low lift configurations SYMBOLS FOUND ON THE OC 2300 CHAIR Protective Earth ground location CAUTION Electrical shock possibility ATTENTION Information provided in manual Alterna...

Страница 6: ... appropriate wall outlet or into the ophthalmic stand 2 Connect the chair control cable to the ophthalmic stand and check the vertical movement of the ophthalmic chair 3 4 Alternate Connections When the Ophthalmic Chair model OC 2300 is not used in conjunction with a Topcon ophthalmic stand use the table below to configure the electrical inputs correctly Stand Cable Wire Code Pin Number Wire Funct...

Страница 7: ...nt of the full power recline ophthalmic chair model OC 2300 is made by pressing the appropriate arrow button on either membrane switch located on the sides of the back or with an optional footswitch The chair may also be controlled by the UP DOWN and AUTORETURN buttons on the exam stand 4 2 Motorized Recline Control The motorized recline buttons located on the sides of the backrest recline and upr...

Страница 8: ...ditional safety feature pressing any button on either chair membrane switch will also terminate any automatic motion 4 6 Swivel Brake Operation To rotate the chair move the swivel brake handle to the released position perpendicular to the floor rotate the chair to the desired position then move the swivel brake handle to the activated position parallel to the floor to secure the chair in position ...

Страница 9: ...e swivel brake handle is in the activated position parallel to the floor an adjustment may be required The swivel brake adjustment screw is located behind the lift assembly on the underside of the metal chair seat If the swivel brake clamping tension is too loose move the swivel brake handle to the released position perpendicular to the floor and turn the adjustment screw with an Allen wrench Cloc...

Страница 10: ...e chair from the power source WARNING Any service and or repairs other than replacement of the two exter nal fuses shall be performed only by a qualified technician The main power fuses are located in the rear of the base assembly in the LH side plastic shroud Unplug the chair from the AC power source Twisting the fuse cap counter clockwise and simultaneously pressing down easily replace the fuses...

Страница 11: ...8 If problems persist call your dealer or Topcon for assistance Topcon Medical Systems Inc 37 West Century Road Paramus NJ 07652 800 223 1130 www topconmedical com ...

Отзывы: