ASSEMBLY
3.1
Tools
Phillips Screw Driver
5/32” and 3/16” Allen Wrench
3.2
Ophthalmic Chair / Stand Position
1)
Place the chair next to the base of the ophthalmic stand or cabinet. Allow it to
come to room temperature.
2)
Remove the red shipping bolt, which secures the chair seat to the base
assembly (Remove the seat upholstery if installed).
WARNING: Red shipping bolt must be removed before power up.
!
3)
Install the upholstery set to the chair (if not already installed) with the
instructions included in the upholstery box.
3.3
Electrical Connections
1)
Connect the chair power cable into the appropriate wall outlet or into the
ophthalmic stand.
2)
Connect the chair control cable to the ophthalmic stand and check the vertical
movement of the ophthalmic chair.
3.4
Alternate Connections
When the Ophthalmic Chair model OC-2300 is not used in conjunction with a
Topcon ophthalmic stand, use the table below to configure the electrical inputs
correctly.
Stand Cable Wire Code
Pin
Number
Wire Function
Wire Color
1
Ground
Green
2
Neutral
White
3
Hot
Black
4
Up
Red
5
Down
Violet
6
Auto-Return
Orange
7
Relay Engage
Brown
8
Safety Switch
Yellow
9
Safety Switch
Blue
WARNING:
If the chair is used with a Topcon IS-800 stand, the plastic base shroud (non-fuse
side) must be removed to make a configuration change on the circuit board. This
will make the IS-800 safety switch feature operable.