Topcon Europe Medical BV
Instruction Manual CC-100XP/ CV-5000PC
19/42
Stereo test
Stereo test
Stereo test
Figure 22, Standard polarization chart selection in CV-5000PC operation window
When the above tests are selected by pressing the related button, the related test will be
displayed in CC-100XP and the related Polaroid filters will be engaged within CV-5000
optical head.
8.1.2 Inverting the display of the Polarized tests within CC-100XP.
In conventional Polaroid testing the inversion of the tests is performed by changing the
Polaroid filters within the CV-5000 Optical Head. In case the CC-100XP is operated the
inversion of the Polaroid tests is performed by the CC-100XP software. There is no physical
change of Polaroid filters within the CV-5000 Optical Head.
For the available Polaroid tests different inversion procedures are available.
8.1.2.1
Polaroid inversion by toggle function of the button
For two tests the toggle function is available to invert the Polaroid display; pressing the
button a second time will invert the display of the Polaroid tests within CC-100XP
Coincidence test (H)
Coincidence test (V)
Figure 23, Polaroid tests with toggle function