1. Before Use
16
5 FACTOR A/B
FACTOR A
: Select when setting a single correction factor.
The correction is set in the function mode.
FACTOR B
: Select when setting the direct correction function, using other instrument
BM-7AC, SC-777 or SR-3. The correction is set in the function mode.
Function Mode
☞
「
3.1 Function Mode
」
Direct Correction Function
☞
「
2.1.4 Reducing Error Among a Plurality of
Measurement Instruments and Performing
Measurement
」
・
This switch can be changed during power ON.
6 NC
This is an empty switch and should always be set to the ON side.
7 NC
This is an empty switch and should always be set to the ON side.
8 PROTECT/WRITE
PROTECT
: Prevents writing to internal memory.
This switch should always be set to the PROTECT side during measurement.
WRITE
: Allows writing to internal memory.
This switch should be set to the WRITE side when changing memory contents.
・
This switch can be changed during power ON.
Request
・
Put the cover on the internal switches/buttons during measurement to prevent a
light from intruding into instrument. Otherwise accurate measurement result
can not be obtained.
Memo
Memo