3. SETTING PROCEDURES
3.1 CORRECTION FACTOR(C.C.F. setting)
Associated mode: Luminance / Illuminance / UV mode
When other than 1.000 is entered, correction factor is active, When 1.000 is entered,
correction factor is inactive.
1
Push the [C.C.F.] switch to enter the Color correction factor setting.
2
Enter a numeric value through the [0
– 9] switch.
3
To shift the digit, push the [SHIFT] switch.
To change the numeric value after shifting the digit, repeat procedure 2.
4
To enter a decimal point, push the
[●] switch.
4
Push the [SET] switch to activate the entered numeric value.