2.5 CALIBRATION FOR PHOSPHORESCENCE
LUMINANCE PREDICTION
Associated mode: Luminance mode
2.5.1 INTRODUCTION
The function of the Calibration for phosphorescence luminance prediction can calibrate the BM-100
for user-specified phosphorescent material. Based on the results of the calibration, the BM-100 can
predict phosphorescence luminance of user-specified material at elapsed time of 20, 60 and 90
minutes by 60 seconds measurement.
☞
‘
2.4 PHOSPHORESCENCE LUMINANCE PREDICTION
’
Exciting user-specified material by three different intensity of illuminance and Measuring
phosphorescence luminance of user-specified material at each intensity of illuminance, and then
calibrating the BM-100 by using Measurement program MT-100.
☞
“MT-100 4.3 Calibration for Phosphorescence luminance prediction”
When you predict phosphorescence luminance of user-specified material after calibration, select
material number 0 (User-specified material) in the material selection menu.
☞
‘
2.4.1 THE MATERIAL SELECTION FOR PHOSPHORESCENCE LUMINANCE PREDICTION
’
User-specified materia
l
Calibration for phosphorescence luminance prediction
Exciting
Pre-measurement
(
5400 seconds
)
Transferring measured data to
MT-100
Measuring three times
Calibrating BM-100 by using
MT-100
キャリブレーション
NO
YES