Check the measurements every day when turning on the device using the instrument provided.
The user of the device must check that the measurements provided by the device are plausible.
It is advisable to visually check all the light sources before examining patients to make sure that they
come on properly.
If the device frequently emits error signals, turn it off and contact technical support to have the device
checked.
In patients with blue eyes, acquisition of pupillometry in mesopic lighting conditions can be
difficult to accomplish. In this case, we suggest acquiring the mesopic data through dynamic
pupillometry.
Contact lenses must not be worn by the patient during data acquisition.
6.9
Cybersecurity
When performing the installation of a new unit the user MUST set his own credentials to prevent
unauthorized physical access to the device.
Make sure the USB devices you intend to connect to the instrument are secured against
malware/viruses.
Patient data on USB devices can become corrupted when inserting into computers for backup or
transfer.
The use of antivirus software on computers is recommended and it is responsibility of the user.
To protect data exported to USB from unauthorized access, use dedicated USB data for storage.
Installation of any unapproved software, including drivers, could degrade the performances of the
instrument and may void the instrument warranty.
Содержание ALADDIN
Страница 29: ...ALADDIN User manual Rev 23 of 08 07 2016 29...
Страница 32: ...ALADDIN User manual Rev 23 of 08 07 2016 32 12 Setting up the instrument 12 1 Connection modes Figure 13...
Страница 42: ...ALADDIN User manual Rev 23 of 08 07 2016 42...
Страница 46: ...ALADDIN User manual Rev 23 of 08 07 2016 46 Figure 26 Figure 27...
Страница 111: ...ALADDIN User manual Rev 23 of 08 07 2016 111 Figure 84...