4.1
EMC table
Emission issues
ALADDIN is intended for use in the electromagnetic environment specified below. The customer or user must ensure
that it is used only in such an environment.
Emission test
Compliance
Electromagnetic environment - guide
RF emissions
CISPR 11
Group 1
ALADDIN uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
neighbouring electronic devices.
RF emissions
CISPR 11
Class B
ALADDIN is suitable for use in all residential buildings and those directly
linked to a low-voltage supply network that supplies residential buildings.
Harmonic Emissions
IEC 61000-3-2
Class A
Compliant
The device can be used in all buildings, including residential buildings and
those connected to the low-voltage mains network that supplies
residential buildings.
Emissions of voltage
fluctuations/flicker
IEC 61000-3-3
Compliant
Immunity issues
ALADDIN is intended for use in the electromagnetic environment specified below. The customer or user must ensure
that it is used only in such an environment.
Immunity issues
Test level
EN 60601-1-2
Compliance level
Electromagnetic environment - guide
Electrostatic discharge
(ESD)
EN 61000-4-2
6kV on contact
8kV in air
6kV on contact
8kV in air
The floors must be wood, cement or
ceramic. If the floors are covered with
synthetic material, relative humidity
must be at least 30%.
Transients/fast electric
Trains EN 61000-4-4
2kV power supply lines
2kV power
supply lines
The power supply should be typical of a
commercial environment or hospital.
Surge
EN 61000-4-5
1kV differential mode
2kV common mode
1kV differential mode
2kV common mode
The power supply should be typical of a
commercial environment or hospital.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
The power supply should be typical of a
commercial environment or hospital. If
the user requires the device to operate
continuously even during a power cut,
we recommend powering the device
with an uninterruptable power supply
(UPS) or a battery.
Содержание ALADDIN
Страница 29: ...ALADDIN User manual Rev 23 of 08 07 2016 29...
Страница 32: ...ALADDIN User manual Rev 23 of 08 07 2016 32 12 Setting up the instrument 12 1 Connection modes Figure 13...
Страница 42: ...ALADDIN User manual Rev 23 of 08 07 2016 42...
Страница 46: ...ALADDIN User manual Rev 23 of 08 07 2016 46 Figure 26 Figure 27...
Страница 111: ...ALADDIN User manual Rev 23 of 08 07 2016 111 Figure 84...