TH2689/89A Operation manual
54
Instruction
:
numeric_value is 4
(
20mA
)、
3
(
2mA
)、
2
(
200uA
)
、
1
(
20uA
)及
0
(
2uA
)
MIN
为
2uA
MAX
为
20mA
6.
:
LCTest
:
CONFigure
:
RANGe
:
AUTO { OFF | ON | 0 | 1}
Function
:
set or query if set on auto switching bin mode
Parameter
:
{OFF | ON | 0 | 1}
Return value
:
{ 0 | 1}
Instruction
:
OFF | 0
select range bin manually
ON | 1
select range bin automatically
7.
:
LCTest
:
CONFigure
:
CHGTime {<numeric_value> | MIN | MAX}
Function
:
set or query the charge time
Parameter
:
{<numeric_value> | MIN | MAX}
Return value
:
numeric_value
Instruction
:
numeric_value
为
0Sec
~
999Sec
MIN is 0Sec
MAX is 999Sec
8.
:
LCTest
:
CONFigure
:
DWELl {<numeric_value> | MIN | MAX}
Function
:
set or query the delay time value of SEQ test mode
Parameter
:
{<numeric_value> | MIN | MAX}
Return value
:
numeric_value
Instruction
:
numeric_value
为
0.2Sec
~
999.0Sec
MIN is 0.2Sec
MAX is 999.0Sec
9.
:
LCTest
:
MEASure
:
STATe
?
Function
:
query the current test status
Parameter
:
none
Return value
:
{CHG | TEST | DCHG}
10.
:
LCTest
:
MEASure
:
FETCh
?
Function
:
query test result
Parameter
:
none
Return value
:
{0
(
OK
)
| 1
(
ERROR
)
}
,
{NO | PASS | HIGH | LOW}
Instruction
:
0
not in TEST status, or in TEST not over the range
1
select a unsuitable range, can not obtain a correct test result
NO
not use comparator to make comparision
PASS | HIGH | LOW
comparator performs compare result
11.
:
LCTest
:
MEASure
:
IR
?
Function
:
query IR value of test result
Parameter
:
none
Return value
:
numeric_value
,
format is <NR3>
,
unit is OHM
12.
:
LCTest
:
MEASure
:
LC
?
Function
:
query LC value of test result
Parameter
:
none
Return value
:
numeric_value
,
format is <NR3>
,
unit is AMP
13.
:
LCTest
:
MEASure
:
VMON
?
Function
:
query the voltage test value
Parameter
:
none