61
6.3.8 RESIister subsystem
The commands in this subsystem are used to configure and control resistance measurement function
and are summarized in Table 6-8.
Table 6-8
RESIister Subsystem Commands Summary
Command Function
Description
Default
:RESistance
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure resistance
Set the test speed and display digits
Query line cycle integration rate
Path to configure measurement range
Select range (
100,1k,10k,100k,1M,10M,100M
)
Query range
Enable or disable auto range
Query auto range
Specify reference (0 to 20e6)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
SLOW, PLAC5
100
ON
0
OFF
: NPLCycles <n>
Command Systac:
:RESistance:NPLCycles <n>
Set the test speed and display digits for
Ω
<n> = SLOW,FAST,PLAC4,PLAC5 Set the test speed and display digits
DEFault SLOW
SLOW Set the test speed as SLOW
FAST Set the test speed as FAST
PLAC4 Set the display digits as 4 1/2
PLAC5 Set the display digits as 5 1/2
Query:
:NPLCycles?
Query the current test speed
Description:
The command is used for the basic measurement functions (except Frequency and Period) and test
speed (display digits).