31
7. Use
keys to setup the parameters and press
to save the setup.
4.2.4 External trigger
BUS trigger: use RS232C or USB bus control to trigger the instrument.
EXT trigger: use START signal pin in HANDLER interface to trigger the instrument.
4.3 Limit Operations
Limit operations set and control the values that determine the HI / IN / LO status of subsequent
measurements. Limits can be applied to all measurement functions except continuity. The limit test is
performed after percent math operations. Unit prefixes are applied before the limit test, for example:
Low Limit = -1.0, High Limit = 1.0
A 150mV reading equals 0.15V (IN).
Low Limit = -1.0, High Limit = 1.0
A 0.6k
Ω
reading equals 600
Ω
(HI)
You can configure the multimeter to beep or not when readings are outside of the limit range.
Application method
Method 1:
1. Press
Shift
+
key to enter the Limit measurement function.
2. Press key and the Hi value is displayed in the secondary display. The initial value is +1.00000;
press
key again, the Lo value is displayed in the secondary display. The initial value is
-1.00000.
3. Press key to enter into the edit interface of Hi or Lo.
4. Use
keys to setup the parameters and press
to save the setup.
Method 2:
1. Press
Shift
→
to enter the menu on the menu level.
2. Use
or
key to select
A: MATH MEU
.
3. Use
or
key to move down to the next level and use
or
key to select 1:HI
LIMIT or 2
:
LO LIMIT.
4. Use
or
key to enter into the parameter setting interface. The default value is HI:
+1.00000; LO: -1.00000.
5. Use
keys to setup the parameters and press
to save the setup.
6. Press
Shift
+
key to enter into the Limit measurement function. The current measurement
value is displayed in the primary display and the current measurement state is displayed in the
secondary display.