6.3.3 VOLTage subsystem
The commands in this subsystem are used to configure and control voltage measurement function and
are summarized in Table 6-3.
Table 6-3
VOLTage Subsystem Commands Summary
Command
Function Description
Default
:VOLTage:DC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n
〉
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure DC voltage
Set integration rate (line cycle; 0.5 to 2)
Query line cycle integration rate
Path to configure measurement range
Select range (0 to 1010)
Query range
Enable or disable auto range
Query auto range
Specify reference (-1010 to 1010)
Enable or disable reference
Query state of reference (0,1)
Use input signal as reference.
Query reference value
1
1000
ON
0
OFF
:VOLTage:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n
〉
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure AC voltage
Set integration rate (line cycles; 0.5 to 2)
Query line cycle integration rate
Path to set measurement range
Select range (0 to 757.5)
Query range
Enable or disable auto range
Query auto range
specify reference (-757.5 to 757.5)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
1
757.5
ON
0
OFF
Speed Commands
: NPLCycles <n>
Command Systac:
:VOLTage:AC:NPLCycles <n>
Set NPLC for ACV
:VOLTage:DC:NPLCycles <n>
Set NPLC for DCV
Command Parameter:
<n> = 0.5 to 2
Set power line cycles per integration
DEFault
1