47
Table 6-1
DISPlay Subsystem Commands Summary
Command
Function Description
:DISPlay
:ENABle <b>
:ENABle?
Enable or disable front panel dispaly
Query state of the display
:ENABle <b>
Command syntax:
:DISPlay:ENABle <b>
Command Parameter:
<b> =
0 or OFF
Disable front panel display
1 or ON
Enable front panel display
Query:
:ENABle? Query state of the display
Description:
This command is used to enable or disable front panel display circuitry. When disabled, the
instrument operates at a higher speed. While disabled, the display is frozen. All front panel controls
except LOCAL are disabled. Normal display operation can be resumed by using:ENABle command
or pressing LOCAL key to enable the display.