53
6.3.4 CURRent subsystem
The commands in this subsystem are used to configure and control current measurement function and
are summarized in Table 6-4.
Table 6-4
CURRent Subsystem Commands Summary
Command
Function Description
Default
:CURRent:DC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure DC current
Set integration rate(line cycles; 0.5 to 2)
Query line cycle integration rate
Path to configure measurement range
Select range (0 to 20)
Query range
Enable or disable auto range
Query auto range
Specify reference (-20 to 20)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
1
20
ON
0
OFF
:CURRent:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure AC voltage
Set integration rate (line cycles; 0.5 to 2)
Query line cycle integration rate
Path to set measurement range
Select range (0 to 20)
Query range
Enable or disable auto range
Query auto range
specify reference (0 to 20)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
1
20
ON
0
OFF
Speed Commands
: NPLCycles <n>
Command Systac:
: CURRent:AC:NPLCycles <n>
Set NPLC for ACI
: CURRent:DC:NPLCycles <n>
Set NPLC for DCI
Command Parameter:
<n> = 0.1 to 10
Set power line cycles per integration