5.5 Replacing consumables
5.5.1 Fuses
■
Disconnect the power cord from the outlet when replacing fuses.
Otherwise, you may get an electric shock, resulting in death or
serious injuries.
■
Use fuses specifically designed for this instrument and replace
two fuses at the same time. Using another type of fuse may result
in failure or fire.
■
When the instrument does not work correctly after fuses have
been replaced, there may be other causes of the problem. Turn off
the instrument immediately and contact our local distributor.
1) Turn power off and disconnect the power plug from the outlet.
2) Disconnect the power cord from the power socket.
3) Insert a flat screwdriver into the groove of the fuse holder under the power socket to
release the lock and remove the holder case.
4) Remove the blown fuse and install new fuses.
5) Reverse the procedure for removing the holder case and return the holder case to
the fuse holder.