1. Confirm that the measurement conditions have been set.
2. Anesthetize the eye with an appropriate topical anesthetic.
3. Instruct the patient to sit back or to lie supine to maximize patient comfort.
1. Select the following settings:
Converted veolcity: 1,640 m/s
Acquisition mode:
Automatic
Measurement mode: Single
Measurement range: 300-1,200
µ
m
Data display method: Actual value
Data presentation mode: Average
2. Immerse the test piece under water.
3. Apply the pachymetry probe perpendicu-
larly in one of the holes in the test piece.
4. Take 10 series of measurements, as
described in section 5.2.2b, and select
the 5 smallest values. The range of these
values should be less than 10
µ
m.
5. Take the average of these 5 values.
6. Repeat steps 3, 4 and 5 for the other
test piece hole.
Expected results (see NOTE below):
Shallower hole (D1): 465
µ
m ± 50
µ
m
Deeper hole (D2): 850
µ
m ± 50
µ
m
N
NO
OT
TE
E:: The expected results are for an
ambient room temperature of 24
degrees C (75 degrees F). Compensate
for higher (lower) temperature by adding
(subtracting), as follows:
D1: 1
µ
m per degree
D2: 2
µ
m per degree
T
Th
he
e b
biio
om
me
ettrryy tte
esstt p
piie
ecce
e iiss u
usse
ed
d o
on
nllyy ffo
orr cch
he
ecck
kiin
ng
g tth
he
e o
op
pe
erra
attiin
ng
g p
pe
err--
ffo
orrm
ma
an
ncce
e o
off tth
he
e iin
nssttrru
um
me
en
ntt.. IItt cca
an
nn
no
ott b
be
e u
usse
ed
d ffo
orr d
de
ette
errm
miin
niin
ng
g tth
he
e
p
prre
ecciissiio
on
n o
off tth
he
e iin
nssttrru
um
me
en
ntt o
orr ffo
orr cca
alliib
brra
attiin
ng
g tth
he
e iin
nssttrru
um
me
en
ntt..
IIff tth
he
e p
pa
attiie
en
ntt iiss rre
ella
ax
xe
ed
d a
an
nd
d cco
oo
op
pe
erra
attiivve
e,, tth
he
e m
me
ea
assu
urre
em
me
en
ntt w
wiillll b
be
e
e
ea
assiie
err a
an
nd
d m
mo
orre
e ssu
ucccce
essssffu
ull.. P
Prriio
orr tto
o tta
ak
kiin
ng
g m
me
ea
assu
urre
em
me
en
nttss,, e
ex
xp
plla
aiin
n
tth
he
e p
pu
urrp
po
osse
e a
an
nd
d m
me
etth
ho
od
d o
off m
me
ea
assu
urre
em
me
en
ntt tto
o tth
he
e p
pa
attiie
en
ntt tto
o rre
ed
du
ucce
e
h
hiiss//h
he
err a
an
nx
xiie
ettyy..
Check the performance of the AL-2000 by using the pachymetry test piece (found
in the box containing the pachymetry probe).
Содержание AL-2000
Страница 9: ...AL 2000 Biometer Pachymeter...
Страница 13: ...AL 2000 Biometer Pachymeter 4...
Страница 41: ...AL 2000 Biometer Pachymeter 32 4 4 3 3 I IO OL L P Po ow we er r C Ca al lc cu ul la at ti io on n...
Страница 89: ...AL 2000 Biometer Pachymeter 80 8 8 M Me em mo or ry y C Ca ar rd d...
Страница 111: ...AL 2000 Biometer Pachymeter...
Страница 118: ...AL 2000 Biometer Pachymeter 109...
Страница 119: ......