![Tohnichi CEM10N3X8D-G-BTA Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/tohnichi/cem10n3x8d-g-bta/cem10n3x8d-g-bta_instruction-manual_1124724019.webp)
19/35
ii. Measurement mode (Default: MODE-M)
Select the mode from “MODE-T” (Tightening mode) or “MODE-M” (Inspection mode).
※
When set to “MODE-M” the double tightening detection angle setting will be skipped. If
there is double tightening detection angle setting and change to “MODE-M” the double
tightening detection angle will be set to “0”.
Select the mode using upper or down arrow keys. Press “MEM” key to save setting and
go to next. Press “MD” key to skip setting and go to next. Press “C” key to go back to
RUN mode (measurement mode).
iii. Unit of torque (Default: N.m)
Select the unit of torque measurement from N.m, kgf.cm, kgf.m, lbf.in or lbf.ft.
Select the unit using upper or down arrow keys. Press “MEM” key to save setting and
go to next. Press “MD” key to skip setting and go to next. Press “C” key to go back to
RUN mode (measurement mode).
iv. Lower limit torque (Default: 0)
※
If the value is set out of tool range the display shows an error "SETERR".
Set the number using down arrow key and move the digit using upper arrow key. Press
“MEM” key to save setting and go to next. Press “MD” key to skip setting and go to next.
Press “C” key to go back to RUN mode (measurement mode).