Instructions for Use
TNI
soft Flow
50
Clinic System
Contraindications
• Do not use the TNI softFlow applicators if you are allergic to silicon.
• The TNI softFlow system is not intended as a life-supporting measure.
• The TNI softFlow system may not be used for invasive ventilation.
• The nasal application of high flow by means of TNI softFlow must not be used if the patient’s upper
respiratory tract is completely obstructed.
• The nasal application of high flow by means of TNI softFlow must not be used if the patient’s upper
respiratory tract has been bypassed using a bypass.
• TNI softFlow system must not be used in patients who have a history of anamnestic seizures or
restless sleep.
• If the patient‘s upper respiratory tract is bypassed, only use the tracheal interface applicator in
connection with an open tracheal interface.
• Do not use the applicator headgear if you have enormous pressure marks from it or in the event of
material incompatibility of the stretch band or applicator headgear.
1.3 System components
TNI
soft Flow
50
PATIENT INTERFACE
APPLICATOR PLUG
DISPLAY
MODE
UP
ENTER
DOWN
CONTROL PANELS
TRACHEAL-APPLICATOR