2.3 Oxygen supply
If supplemental oxygen is required, an external, medically approved oxygen source can be connected to the
TNI softFlow using the lateral oxygen inlet port.
The oxygen inlet port is located
on the left side of the device
casing.
If no oxygen supply is needed,
the oxygen inlet port must be
kept sealed by the protective
cap.
Connect an external oxygen source
to the oxygen inlet port of the
device using a dedicated oxygen
tube.
NOTE
• Please follow the instructions for use of the external oxygen source closely. If you have any
questions concerning the use of the oxygen source, please contact your oxygen vendor or our
hotline [see chapter 7. Service/User assistance information].
•
Incorrect connection of the oxygen source may lead to inefficient oxygen therapy. Ensure a
stable connection.
WARNING
• Secure the oxygen source against falling over, to prevent damage and injury.
•
Smoking and open fire are strictly forbidden when using supplemental oxygen due to the risk
of explosion.
• Do not operate the device in closed rooms producing or using anesthetics and/or nitrous
oxide.
•
Keep the oxygen valves free of oil, grease or any flammable liquids.