Instructions for Use
TNI
soft Flow
50
Clinic System
4.3 Cleaning and replacement cycles
The following cleaning and replacement cycles must be followed strictly. Between patients, single use com-
ponents must be replaced. If necessary, carry out a manual cleaning, for example if superficial dirt is visible
(see chapter 4.1.1 Manual cleaning).
Component
Cleaning cycle
Cleaning method
Replacement cycle Usage
TNI
soft Flow
50
daily
manual cleaning
wipe disinfection
-
reusable
Clinic system
Applicators clinic series
daily
manual cleaning
wipe disinfection
360 hours
single use
Humidifier rack clinic
daily
wipe disinfection
-
reusable
Humidification chamber auto-fill
-
weekly
single use
Air Bridge Humidifier Clinic
daily
wipe disinfection
weekly
single use
Clear-Guard 3 Breathing Filter
-
-
daily
single use
Homecare system
Applicators Homecare series
daily
manual cleaning
wipe disinfection
720 hours
single use
Water Tank Humidifier Homecare
daily
manual cleaning
once a year
single use
Cyclone Humidifier Homecare
daily
manual cleaning
once a year
single use
Lid Humidifier Homecare
daily
manual cleaning
once a year
single use
Miscellaneous
Dust filterr
weekly
rinsing
every 3 months
single use
Water in water tank
-
-
daily
Applicator Headgear
weekly
manual cleaning
wipe disinfection
once a year
single use
Stretchband Applicator Headgear
weekly
manual cleaning
-
-
Applicator Clip
single use