Tisch Environmental
PUF+ Operator’s Manual
47
!,2,3,04-20-07 10:33:00,21.6,26.6,760,51.8,6.24,260,259,0
!,2,3,04-20-07 10:34:00,21.7,26.6,760,51.8,6.22,260,259,0
!,2,3,04-20-07 10:35:00,21.7,26.6,760,51.8,6.23,260,259,0
!,2,4,"End","Interval Log",04-20-07 12:34:47,"PUF+","100-0004"
DATA.CSV: Power Log Example
The Power Log contains information on when the unit started and stopped. This is useful in
identifying when power failures occurred and how long they lasted. The last 64 power
failures are recorded. This can be important in determining whether a sample can be
considered valid if a power failure occurred during the sampling period. The “Source” field
is intended to help technical support track down any power related problems with the unit
and does not contain any data relating to the actual sampling taken place.
!,3,1,"Begin","Power Log",04-20-07 12:34:47,"PUF+","100-0004"
!,3,2,"Off","On","Source"
!,3,3,04-20-07 10:19:35,04-20-07 10:19:38,130
!,3,3,04-20-07 11:48:50,04-20-07 12:02:50,16
!,3,4,"End","Power Log",04-20-07 12:34:47,"PUF+","100-0004"
DATA.CSV: Calibration Log Example
The “Calibration Log contains a list of the times, sensors, number of calibration points used
in the calibration, the calibration gain (A1) and the calibration offset (A0) of each sensor.
The last 64 calibrations are recorded. This information can be used to track how the
calibration of the unit changes over time, help identify potential problem before they occur,
and identify which users may need additional training in calibration a particular sensor.
!,4,1,"Begin","Calibration Log",04-20-07 12:34:47,"PUF+","100-0004"
!,4,2,"Time","Sensor","Pts","A1","A0"
!,4,3,03-02-07 20:39:29,"Tcjc",1,13.771930,0.000000
!,4,3,03-02-07 20:41:47,"Tamb",2,0.299843,-11935.423828
!,4,3,03-02-07 20:46:11,"Pamb",4,30.914890,-13626.343750
!,4,3,03-02-07 20:48:59,"Pdif",4,3.201837,-1770.477539
!,4,3,03-02-07 20:52:18,"Pcal",4,0.854666,-284.707397
!,4,3,03-02-07 20:55:23,"Pcal",5,0.857291,-287.563842
!,4,3,04-20-07 09:19:46,"Pcal",1,0.857291,-281.191467
!,4,3,04-20-07 11:48:15,"Qstd",5,32.117942,-1.297638
!,4,4,"End","Calibration Log",04-20-07 12:34:48,"PUF+","100-0004"
Содержание PUF+
Страница 1: ...PUF Operator s Manual Date 3 18 2008...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 9: ...Tisch Environmental PUF Operator s Manual 5...
Страница 10: ......
Страница 30: ......
Страница 32: ......
Страница 34: ......