Tisch Environmental
PUF+ Operator’s Manual
11
STEP BY STEP GUIDE
This section of the manual serves as a quick guide providing step by step instructions to some
of the most important features of the PUF+. For additional details, refer to the relevant
section of this manual.
Step by Step: Setting the PUF+ System for the First Time
The following steps provide instructions for unpacking and setting the unit up for the first
time.
1. Unpack unit, take PUF module out of box, unwrap exhaust hose and find piece of
calibration tubing (1 per order).
2. Install brushless motor onto ring under Venturi (make sure gasket is in between motor
and ring).
3. Install Lid using the included hardware.
4. Install exhaust hose down wind of unit.
5. Insert module into quick disconnect, and take off module cover. Normally, this is
where the filter media is located. However, in preparation for the initial flow
calibration there should be no media installed, only the glass holder is required.
6. Adjust the yellow thermocouple wire at the top of unit so that is located space that air
will enter the unit will sampling, approximately 1” above top of shelter (pull up).
7. Plug the male power cord into AC Power. The system will power up. It is VERY
important to apply the correct AC voltage (110 or 220 depending on the system
ordered).
Applying an incorrect AC Voltage can result in incorrect operation and
damage to the system.
8. After, 5 seconds the main screen will appear.
9. Verify that your date and time are correct. If incorrect, follow the step by step
instructions for setting the date and time in this section of the manual.
10. The desired operating flow type must be set in accordance with the sampling
protocols. Follow the step by step instructions for Setting the Flow Type in this
section of the manual.
11. The desired operating flow rate must be set in accordance with the sampling
protocols. Follow the step by step instructions for Setting the Flow Rate in this
section of the manual.
12. After allowing the unit to warm up for 30 minutes, perform the step by step
instructions for Auto Calibrating the Flow.
13. The unit should now be ready for normal operation.