Altitude Control Mode (Fig 7.0): Negative pressure at Ps and Pt

 

Air Speed Indicator/ Altimeter 

(Control Equalizing Valve for airspeed)

 

 
       

1.

 

Switch to Altitude Control position  

2.

 

Connect both Ps and Pt via pressure hose to the U.U.T/D.U.T 

3.

 

Set both Isolator valve to OPEN position 

4.

 

Set Relief valve to OPEN position (10 inHg differential pressure protection) 

5.

 

Close the Equalizing valve by turning knob fully clock-wise, 

do not over torque

, hand tight is good 

enough. 

6.

 

Close both Pressure and Vacuum Control metering valve by turning knob fully clock-wise, 

do not 

over torque

, hand tight is good enough. 

7.

 

Switch OFF VENT valve 

8.

 

Switch on the internal Vacuum Pump. 

9.

 

Open the Vacuum Control metering valve slowly anti-clock-wise to increase both knots and feet 
reading. Control Equalizing valve for various airspeed. 

10.

 

Switch OFF the internal vacuum pump and CLOSE Vacuum Control metering valve when readings 
reaches set point. 

11.

 

If require, CLOSE both the isolator valves. 

12.

 

To VENT the system, 

switch the x2 isolator valves to OPEN position, switch ON the VENT valve, 

open the Equalizing valve slowly to equalize the pressure between Pt and Ps. When both Pt and Ps 
pressure equalize, open the Pressure Control metering valve slowly anti-clock-wise to vent vacuum 
pressure. 
 

 

 

Fig 7.0 

Содержание EPC300A

Страница 1: ...Electric Pressure Vacuum Tester Model EPC300A User Reference Manual...

Страница 2: ...as a calibration reference or in any application where high accuracy pressure measurement is required 1 1 Customer Service Corporate Office www tiscalibrator com e mail sales tisinstruments com Tel 65...

Страница 3: ...y damage that may cause a safety concern If in doubt please return the unit for evaluation to TIS Instruments Please refer to the Customer Service Section for contact information Caution To avoid poss...

Страница 4: ...4 EPC300A Front Panel Layout...

Страница 5: ...OWN position To switch OFF the unit flip the ON OFF switch to UP position Upon power on display shown the main menu refer to Fig 1 0 Main menu Fig 1 0 1 Engineering Units Selection By pressing the Uni...

Страница 6: ...mode only Line 2 Ps feet a Vent the system by switching the x2 isolator valves to OPEN position anti clockwise about 3 5 turns both the Pressure and Vacuum control knob switch on the VENT valve b Sel...

Страница 7: ...trol metering valve by turning knob fully clock wise do not over torque hand tight is good enough 7 Switch OFF VENT valve 8 Switch on the internal Pressure Pump 9 Open the Pressure Control metering va...

Страница 8: ...etering valve by turning knob fully clock wise do not over torque hand tight is good enough 7 Switch OFF VENT valve 8 Switch on the internal Vacuum Pump 9 Open the Vacuum Control metering valve slowly...

Страница 9: ...ght is good enough 7 Switch OFF VENT valve 8 Switch on the internal Vacuum Pump 9 Open the Vacuum Control metering valve slowly anti clock wise to increase vacuum level Control Equalizing valve for Pt...

Страница 10: ...re and Vacuum Control metering valve by turning knob fully clock wise do not over torque hand tight is good enough 8 Switch OFF VENT valve 9 Switch on the internal Vacuum Pump 10 Open the Vacuum Contr...

Страница 11: ...ully clock wise do not over torque hand tight is good enough 7 Switch OFF VENT valve 8 Switch on the internal Vacuum Pump 9 Open the Vacuum Control metering valve slowly anti clock wise to increase bo...

Страница 12: ...ct the normally open or normally close U U T D U T wires onto the front panel marked Switch Input Non voltage input only 2 Switch status reading OPEN or CLOSE is display at the top left hand corner No...

Страница 13: ...e 128x4 dots Display Information Line1 Pt and Line2 Ps Line 1 and Line 2 display both Feet reading Pressure Generation Built in electric pressure pump self generate 30 psig Vacuum Generation Built in...

Страница 14: ...14 Pneumatic Circuit Diagram Note Pt address 1 Ps address 2 Baro address 3...

Страница 15: ...uum standards able to produce and indicate pressures from vacuum to the full scale range of the unit under test In order to maintain the specified accuracy of the EPC300A reference standards should ha...

Страница 16: ...ion press SAVE key follow by Ret key Calibration span factor has now saved in EEprom of the CPU Screen proceed to LINE 2 calibration Fig 9 1 Fig 8 0 Fig 8 1 Fig 8 2 Entering Calibration Mode Ps Sensor...

Страница 17: ...he span factor by pressing the UP key change the value flashing and Shift key to shift digit left c After completion press SAVE key follow by Ret key Calibration span factor has now saved in EEprom of...

Страница 18: ...requires service contact our customer service department for a Return Materials Authorization number This number must be indicated on the return package Be sure to package the product well we are not...

Страница 19: ...19 This Page intentionally left blank...

Страница 20: ...20 www tiscalibrator com e mail sales tisinstruments com Tel 65 6779 9272 Fax 65 6873 8262 8 Tradehub 21 06 16 Boon Lay Way Singapore 609964...

Отзывы: