Implementation of the provision Acid capacity K
S4.3
with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with
10 ml
sample
.
Close vial(s).
Sample
Place
sample vial
in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place
sample vial
in the
sample chamber. • Pay
attention to the positioning.
Test
Press the
TEST
(XD:
START
)
button.
The result in Acid Capacity K
S4.3
appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20
EN Handbook of Methods 01/20
How to use
EN Handbook of Methods
5
EN