Ammonia PP / M62
EN Method Reference Book 1.0
12
EN
Close vial(s).
Dissolve the contents by
shaking.
Press the
ENTER
button.
9
4
5
6
7
8
10
11 12
1
2
3
15 min
Blank
Zero
Wait for
15 minute(s)
reaction time
.
Place
blank
in the sample
chamber. • Pay attention to
the positioning.
Press the
ZERO
button.
Sample
Test
Remove the vial from the
sample chamber.
Place
sample vial
in the
sample chamber. • Pay
attention to the positioning.
Press the
TEST
(XD:
START
)button.
The result in mg/l Ammonium appears on the display.