![Time Electronics 7085A Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/time-electronics/7085a/7085a_user-manual_1116926018.webp)
Time Electronics
User Manual
7085A Temperature Distribution and 10 Channel Scanner Module
v1.1
7085A Temperature Distribution and 10 Channel Scanner Module
Page 18 of 21
3.2
Thermocouple Simulation (7051 Module only)
When using a 7051 to source simulated thermocouples the 7085A provides a suitable 4mm
to thermocouple interconnection along with an improved cold junction reference.
Connections
Each 7085A thermocouple socket is connected in parallel
with each red and black 4mm terminal. Connect the 7051
source main terminals to the 4mm sockets and, using a
thermocouple compensating lead, connect the
thermocouple socket to the UUT.
The 7085 is not required to perform any switching as there is a direct connection between
the 7051 & 7085A.
7051 Source App
Select thermocouple from the
7051 source application.
From here select the 7085A as
the CJ reference.
EasyCal
This selection is also used in EasyCal for any thermocouple source tests performed by the
7051. If a manual CJ reference is entered this value will also be used.