4.6 Testing the mask during use
1. Perform this procedure before using the mask,
hence before to enter the environment with the hazardous mixtures
2. This procedure is to be performed as the mask is on the head of the user, after
the procedures: 4.4 Calibrate the blower on page 31, 4.3 Put the Tiki medical
mask on, on page 30.
3. If the mask is tight to the face and calibrated, test the warning system by
performing the following:
a. With one hand pull the mask forward as indicated on figure 12.
Do this to create a leak.
b. Verify that the alarm is activated as to sub-paragraph:
2.3.4 Filter warning or leak on page 17
c. If the alarm was not activated, verify that the batteries are fully charged and try
calibrating again.
d. If the alarm was activated, position the mask again and tighten the straps of the
harness again
4. Note that the test before any use, was to be performed before any use!