@TI-02  USER  MANUAL 

 

SPECIFICATIONS 
 

Accuracy: 

Class III 

Nonlinear: 

≤0.01%F.S 

Excitation voltage:    DC 5V,

up to 4 load cells 350 Ω 

Large display connector:    Sampling serial output way 
Sample rate: 20 times per second (can select) 

 
POWER 

 
Input: 120~240V     
Output: 9V/500mA   
Rechargeable battery: 6V/1.3AH     
 

BEFORE USING   

 
1)  Place this product on a firm and smooth place, don´t place it in vibration or shaking, 

use bench for use on four only adjust foot, adjust the balance using the bubble level. 

2)  Use independent source, avoid other electrical disturbance. 
3)  Don´t put any object on the platter when turn on the balance. 
4)  Please, turn on 2-3 minutes before using. 
5)  Avoid temperature change too large and air flow strenuous sites. 
6)  Don´t overload the balance, don´t exceed the maximal capacity. 
 

SYSTEM POWER CONSUMPTION 

 
Main system power consumption:  about 12mA 
Main system power consumption (with backlight):    about 36mA 
Main system power consumption (with backlight and RS-232):  about 48mA 
Battery life:    none backlight, about 320 hours. 
 

DISPLAY DESCRIPTION 

TI-02 Model: 

 

 
 
 
 
 
 
 
 
 

 
 
 

Содержание TI-02

Страница 1: ...WEIGHING INDICATOR user manual Please read this manual carefully before operation Important safety info Operating condition and other attentions Guarantee info Model TI 02 ...

Страница 2: ...rruptive gas or liquid flush water Don t touch the keypad with the Don t open the indicator by yourself and items with sharp end please call the service people or our distributor when you get the problem with it Please recharge the scale on time to assure the health of the battery and replace the battery if it runs out of life The replaced battery must be handled correctly Important Safety And Ope...

Страница 3: ...r setting 6 UF 1 A D count 6 UF 2 High Low limits setting 6 UF 3 Auto power off 7 UF 4 Backlight setting 7 UF 5 Hold function 7 UF 6 RS 232 output 8 UF 7 Speed setting 10 UF 8 Zero tracking 10 UF 9 G Value setting 11 UF 10 Filt setting 11 UF 12 Beep setting 11 Weight Calibrations 12 ECF 1 Zero and Span Calibration 12 ECF 2 Zero Calibration 12 ECF 3 Span Calibration 12 Guarantee 13 INDEX ...

Страница 4: ...ly adjust foot adjust the balance using the bubble level 2 Use independent source avoid other electrical disturbance 3 Don t put any object on the platter when turn on the balance 4 Please turn on 2 3 minutes before using 5 Avoid temperature change too large and air flow strenuous sites 6 Don t overload the balance don t exceed the maximal capacity SYSTEM POWER CONSUMPTION Main system power consum...

Страница 5: ...net weight when the balance is on tare status All other keys will be disabled when gross weight is activated Function 2 To increase values upward in setup mode Function 1 Key of confirmation in setup mode Function 2 Manually transmitting data through RS232 to computer or printer Function 3 To accumulate weights the max of the weight times is 9999 Function 1 To recall accumulations Function 2 Long ...

Страница 6: ... exit back to menu UF 1 3 To move to next parameter press the key 4 To exit and return to normal weighing press the key UF 2 High Low limits setting 1 Press the key to enter 2 The display will show 00 000L set low limit 3 Use the keys and to move cursor and press the to select number 4 Press the key to confirm 5 The display will show 00 000H set high limit 6 Use the keys and to move cursor and pre...

Страница 7: ...Modes AoFF00 Auto turn off disable AoFF01 The balance will automatically turn off after 1 minute of non use This time can be set up to 99 minutes 1 Press the key to access to auto power configuration 2 Use the keys and to move cursor and press the to select number 3 Press the key to confirm UF 4 Backlight setting Modes lit a Automatic lit on Backlight on litoFF Backlight off 1 Press the key to acc...

Страница 8: ... number this can set the range from 001 100 units of the animal hold 3 Press the key will show TiME 8 4 Use the key to select 1 2 4 8 16 32 or 64 times within the hold range 5 Press the key to confirm UF 6 RS 232 Output 1 Press the key to enter 2 The display will show 232 1 3 Press the key to select the modes 4 Press the key to confirm and will show b 9600 5 Use the key to select the baud rate 6 P...

Страница 9: ...ate 38400 Remarks Format 1 output examples Format 2 output examples Format 3 output examples Format 4 output examples Format 3 and format 4 if you want to print the total weight press the key twice and the accumulative weight will be removed Communication Protocol UART signal of EIA RS232 C Format 1 Baud rate 9600 2 Data bits 8 bits 3 Parity bits None 4 Stop bits 1 bit Format 1 232 1 3 HEAD1 2 BYT...

Страница 10: ... 2 1 1 2 3 4 5 6 7 8 1 2 3 1 2 HEAD1 HEAD2 DATA UNIT CR LF UF 7 Speed setting Modes spEED1 Standard speed spEED2 High speed spEED3 Low speed 1 Press the key to enter 2 Press the key to select the desired mode 3 Press the key to confirm UF 8 Zero tracking Modes Zp 0 OFF Zp 1 One division tracking at zero Zp 2 Two division tracking at zero Zp 3 Three division tracking at zero Zp 4 Four division trac...

Страница 11: ...e new G value 3 Press the key to confirm UF 10 Filt setting Modes FiLt 1 Low FiLt 2 FiLt 3 FiLt 4 High 1 Press the key to enter 2 Press the key to select the desired mode 3 Press the key to confirm UF 12 Beep setting Modes yS oN ON yS oFF OFF 1 Press the key to enter 2 Press the key to select the desired mode 3 Press the key to confirm Weight Calibrations 1 In weight mode press the and keys the di...

Страница 12: ...ghing mode automatically ECF 2 Zero Calibration 1 Press the key to enter display will show cALZ 2 Press the key to calibrate zero point 3 The scale will return back to weighing mode automatically ECF 3 Span Calibration 1 Press the key to enter display will show the calibration weight 00 6 000 2 Use the keys and to select the digit press key to input the weight value 3 Put the calibration weight on...

Страница 13: ... cost and the spare parts necessary for the reparation of the scale This guarantee does not cover the failures caused by an inappropriate use or overcharge The guarantee does not cover the freight cost transport necessary to repair the scale Officially Distributed By ...