@TI-02 USER MANUAL
12
2. Press the or key to select
ECF-1
,
ECF-2
or
ECF-3
.
ECF-1
Zero and Span Calibration
1. Press the key to enter, display will show
cALZ
.
2. Press the key to calibrate zero point.
3. The display will show the calibration weight
00.6.000
.
4. Use the keys and to select the digit,press key to input the
weight value.
5. Put the calibration weight on the platter and press the key to calibrate.
The scale will return back to weighing mode automatically.
ECF-2
Zero Calibration
1. Press the key to enter, display will show
cALZ
.
2. Press the key to calibrate zero point.
3. The scale will return back to weighing mode automatically.
ECF-3
Span Calibration
1. Press the key to enter, display will show the calibration weight
00.6.000
.
2. Use the keys and to select the digit,press key to input the
weight value.
3. Put the calibration weight on the platter and press the key to calibrate.
The scale will return back to weighing mode automatically.
GUARANTEE
This scale is guaranteed for one year from the delivery date. The guarantee covers any
fabrication defect of the material.