18
Measure-
ment
Measuring
ranges
(auto-ranging)
Resolution Frequency
range/
measuring
Intrinsic
error
U-
U~
TRMS
U~¹)
R
0,10 V ... 8,99 V
9,0 V ... 99,9 V
100 V ... 1500 V
0,01 V
0,1 V
1 V
_
±1,5 %
+3 digits
1,0 V ... 99,9 V
100 V ... 1200 V
0,1 V
1 V
15 Hz ...
1,8 kHz
±1,5 %
+3 digits
15 V ... 99,9 V
100 V ... 499 V
›1,8 Hz ...
... 10 kHz
±15 %
+3 digits
0 ... 49 kΩ
50 ... 1999 kΩ
1 kΩ
5 µA
1 µA
±5 %
+2digits
¹) effective value; sinus
7. Characteristic values
500 V ... 1200 V
›1,8 Hz ...
... 4 kHz
Nominal voltage range:
24* ... 1000 V AC/1500 V DC
*Auto on from 24 V
Nominal frequency range: 15 Hz ... 10 kHz < 500 V,
15 Hz ... 4 kHz < 500 V
Input resistance:
311 kΩ
Current (Peak value I
s
):
3,2 mA at 1000 V
On-period:
2 minutes
Display:
5 LEDs for voltage,
continuity, phase and
phase sequence
LCD digital display
7-segment-figure,
2 lines
0 ... 1999 digit
background lightning
3 measurment/s
Power supply:
9 V block battery
IEC 6F22 zinc-carbon or
IEC 6LR61 alkaline
manganese or corre
sponding accumulator,
multi-stage display
of battery status
Surge voltage category: CAT IV
Impulse withstand voltage: >12 kV (1,2/50 µs)
Test voltage:
6 kV
EMV-requirements:
DIN-EN 61326
Operating temperature: -10…+55°C
Casing:
impact resistant,
dustproof plastic casing
with unbreakable display
cover
Protection category:
IP 65
Interconnection line:
rubber-insulated flexible
cable H07RNF, 1m
Dimensions:
test probe with display
part
240 x 62 x 39 mm
Weigth:
270 g (incl. battery)