27
Connect the annular electrode and the upper
socket of the ΩMegaSafe ISO 1Ex with the
measuring line
Place the test object on a good insulating
underground.
Weight the annular electrode with about 2,5 kg
e. g. standard weight / electrode 73012
Test the surface resitance with the ΩMegaSafe
ISO 1Ex
5. Measuring arrangements
For some measurements the following optional
equipment is needed:
- One/two tripod electrodes, art.-no. 73013
(EN 1081)
- An annular electrode, art.-no. 73011
(EN 61340-5-1/61340-2-3)
- An electrode for surface resistance and volume
resistivity in accordance with at once standard
weight of 2,5 kg, art.-no. 73012
(EN 61340-4-1/61340-2-3 )
- Measuring lines 4 mm unisulated socket,
art. no. 73009
5.1
Surface resistance with annular electrode and
standard weight of 2,5 kg (art.-no. 73012):
e. g. for antistatic surfaces like antstatic packaging
Connect the tripod electrode and the upper
socket of the ΩMegaSafe ISO 1Ex with the
measuring line
Place the test object on a good insulating
underground.
Connect the second tripod electrode and the
lower socket of the ΩMegaSafe ISO 1Ex with the
measuring line
Place the tripod electrodes at a distance of
100 mm on the test object and weight both with
approx. 30 kg each.
Test the surface resistance with the
ΩMegaSafe ISO 1Ex.
5.2
Surface resistance with two tripod electrodes:
e. g. for large antistatic surfaces
test object
annular
electrode
standard weight 2,5 kg
test object
insulating
underground
30 kg
30 kg
100 mm