34
vd81-201113
5 Configuration
To switch the VD81 into Configuration Mode:
with the instrument switched-off hold the
Mode-Key pressed for approx. 5 seconds,
until the display shows ‘’rAtE’’.
5.1 Logging Rate
To set the logging rate of your VD81 and thereby activate the data logging
functionality, switch the unit into Configuration Mode. For this the instrument
must be switched-off. Hold the Mode-Key pressed then, until the display shows
‘’rAtE’’:
After additional 5s the current rate setting for internal data logging is displayed
and can now be adjusted by means of the Mode-Key:
off, HiLo, 0,05s / 0,1s / 1,0s / 2,0s / 10s / 1min / 10min und trig.
"off" : data logging is disabled.
"HiLo" means that only minimum and maxi-
mum pressure is stored. These values can be
displayed but cannot be transmitted to a PC!
Save measurements every 0.1 seconds.
…
"trig" means that new measurements are
saved only if the new value differs more than
2 digits from the one that was stored last (e.g.
23
→
25). Data volume is reduced this way
and optimum memory utilization achieved.
When data logging is active minimum and maximum pressure are recorded
simultaneously.
MODE
5s
5s
5s
MODE
MODE
MODE
MODE
s