36
vd81-201113
Afterwards the unit switches to Auto-Off Mode.
Adjustment on atmosphere pressure is possible only if the ac-
tual pressure is above 800 mbar. Otherwise adjustment is de-
nied and the error message "Err" displayed.
Adjustment on Zero Pressure
For adjustment on zero pressure the actual pressure inside the
sensor has to be less than 0,1 mbar !
The pressure reading must be less than 100 mbar, otherwise
adjustment is denied and the error message "Err" displayed.
Switch the unit to Configuration Mode like described above and press Mode-Key
several times, until "CAL" is displayed.
After 5 more seconds the display shows:
After 5 more seconds the display shows:
Press Mode-Key for adjustment. During the
adjustment procedure the display shows
"CALI".
When the adjustment procedure is finished, the unit switches to Auto-Off Mode.
5.3 Pressure Units
To set the displayed pressure unit, switch the VD81 into Configuration Mode. For
this the instrument must be switched-off. Hold the Mode-Key pressed then, until
the display shows "rAtE".
Then press Mode-Key several times, until the display shows "unit":
5s
5s
MODE
MODE