Part 1. Safety
All statements regarding safety of operation and technical data in this instruction manual will only apply
when the unit is operated correctly. Handle all optics with care. Gloves are recommended for optic
installations. Take care when tightening optics, as overtightening optics may cause the optic to break or
shatter.
Part 2. Assembly Instructions
Step 1
Carefully unpack the Olympus MDFM-MF2
Filter Cube assembly. Take note where the
emitter and exciter sides of the cube are
(see Figure 1).
Figure 1. MDFM-MF2 Filter Cube Assembly
Note the orientation of the retainer rings. The flat side of both retaining rings should always face away
from the cube. The side of the retaining ring with the mounting lip should face inwards towards the
filter and dichroic mirror.
Step 2
Remove the two screws holding the filter cube
together. Carefully separate the two filter cube
halves.
Figure 2. Disassembling the filter cube
Exciter Side
Emitter Side
Note the
orientation of
retainer rings
Screw
Locations