Part 7.
Appendix
7.1.
Setting the Line Select Switch
The Swept Source will operate from AC line voltages ranging from 100-240 VAC at 50-60Hz. There is no need to
configure the unit for specific line voltages.
7.2.
Changing the Input Fuses
If for some reason you need to replace an open fuse in the Swept Laser Source, you must perform the following
procedure:
•
Remove the AC input cable that may be connected to the unit.
•
Slide open the cover of the fuse holder located at the rear panel of the Swept Laser Source as shown in
figure 48. Remove the existing fuse and install the appropriate replacement fuse for Swept Laser Source.
Use only is 1A 250VAC Type T 5x20mm style fuse
(IEC 60127-2/III, Low Breaking Capacity, slow blow)
•
Slide the fuse cover closed.