SA201, SA201-EC
Chapter 2: Description
6679-D02
Page 8
Figure 4 SA201 Rear Panel Connections
Ground Plug
This ground plug is for use as a general-purpose ground connection. It is
connected directly to the earth ground connection of the input power plug.
AC Input Connector
This is the line voltage input connection. IMPORTANT: The unit is configured for
100/115 VAC, 50 – 60 Hz from the factory. To operate at 230 VAC see page 11.
PD Amplifier Input BNC
This input BNC is used to interface the photodetector, provided with the SA200
scanning heads, to the amplifier circuit. The photodiode amplifier is configured to
operate with the Thorlabs supplied photo detectors; however it is possible to
operate user supplied photo detectors. To do so, the BNC center contact must be
connected to the photo detector cathode and the BNC shell must be connected
to the photodiode anode (unbiased operation). If a biased detector is to be used
the BNC shell must be connected to the bias ground and the bias voltage must
be negative for the circuit to operate properly.
PD Amplifier Output BNC
This BNC is the amplifier output and may be connected directly to an
oscilloscope to view the cavity spectrum. The amplifier gain will be set using the
front panel ‘DETECTOR’ control knob. The amplifier output includes a 50
Ω
series resistor to minimize noise when operating with a 50
Ω
coax cable. For best
Ground Plug
AC Input Connector
PD Amplifier Input BNC
PD Amplifier Output BNC
Voltage Selector