Chapter 3: Operation
6679-D02
Page 12
3 Locate the fuse box between the input line voltage connector and the
transformer.
4
Remove the fuse cover and slide the old fuse out.
5
Install the new fuse into the fuse cover and place back into the fuse box.
(250 mA for 100/115 VAC and 125 mA for 230 VAC)
6
Replace the enclosure cover and secure with the enclosure screws.
Selecting the Line Voltage
1. Replace the line fuse as described above.
2. Locate the voltage selector switch on the rear panel. See page 9.
3. Switch to the appropriate line voltage.
4. Install the appropriate line cord.
1.1.3. Cleaning
The SA201 should only be cleaned with a soft cloth and a mild soap detergent or
isopropyl alcohol. Do not use a solvent-based cleaner.
Fuse with
Fuse Cover on