Spectrum Analyzer Controller
Page 7
Rev D, February 26, 2013
Figure 2 Trigger Logic
Output BNC
The output BNC is used to drive the SA200 scanning piezos from 1 to 45 V. The
output is capable of driving 0.6 µF piezo loads at a ramp rate of 1 ms over the full
voltage range. The output current is internally limited to prevent damage to the
output drive. Note: the output performance specifications assume a Thorlabs
Fabry Perot Interferometer module is connected.