24
HA0302T Rev A Feb 2016
Chapter 4
4.7 Jogging
During PC operation, the motor actuators are jogged using the GUI panel arrow keys.
There are two jogging modes available, ‘Single Step’ and ‘Continuous’. In ‘Single
Step’ mode, the motor moves by the ste p size specifie d in the Ste p Distance
parameter each time the key is pressed. In ‘Continuous’ mode, the motor actuator will
accelerate and move at the jog velocity while the button is held down.
1) On the GUI panel, click the ‘Settings’ button to display the Settings panel.
Fig. 4.5 Settings Panel - Move/Jogs Tab
2) Select the Move/Jogs tab as shown in Fig. 4.5.
3) In the ‘Jogs’ field, enter parameters as follows: (see Section 5.3.1. for details of
these parameters).
‘Max. Vel’ - ‘50’
‘Accn/Dec’ - ‘50’
Operating Modes
‘Jogging’ - ‘Single Step’
‘Stopping’ - ‘Profiled’
‘Step Distance’ - ‘0.5’
4) Click ‘OK’ to save the settings and close the window.
5) Click the Jog Arrows on the GUI panel to jog the motor. Notice that the p osition
display increments 0.5 mm every time the button is clicked.