The PRO8000 (-4) / PRO800 “ELCH” – macro functions 55
_________________________________________________________________________________
_
_____________________________________________________________________________________________________
© 2012 Thorlabs
4.7.5 Which values do you want to measure?
First select the desired module by means of the
"
:SLOT n
"
command.
Then enter the value to be measured (e.g. ILD, VLD etc.) with the compound
"
:MEAS
.
"
e.g.
:ILD:MEAS n
n stands for the position of the measured value in the output string and can have
values from 1 to 8.
You can not choose fixed parameters (e.g. a hardware limit ILIM etc. or values that
must be calculated from others). If you try to do, you will get an error message ( 310:
ELCH set value initialization not complete) when starting the ELCH measurement.
If you define less values to be measured than defined with
:ELCH:MEAS n
, the
error message “ 311: ELCH read value(s) initialization not complete” is given when
starting ELCH.
The variable parameter always shows up in the first position of the output string and
must not be programmed explicitly.
In the answer string the values are separated by commas.