":TYPE:SUB?"
Queries a possible sub type of a module. E.g.: CW,
EA, DIR, LS, SLED for the optical source modules,
depending on the modulation type.
":TYPE:TXT?"
Module ID in text form.
":TYPE:OPT?"
Returns 10 figures (0...255) depending on the
installed options (mainly for service purposes)
":TYPE:SN?"
Query PRO8 serial number
4.3.5 Module specific commands and queries
See operating manual of the corresponding module
4.4 Error messages
Devices following the IEEE 488.2 standard provide an error queue storing errors one
by one. Every query
":SYST:ERR?"
will fetch one error from the error queue.
Repeated use of
":SYST:ERR?"
is necessary until the error queue is empty.
If the queue is empty, the error message
[0, "No error"]
is sent to the PC.
Error messages are organized in groups:
Error number
Error type
0
no errors
100 ... 199
command errors
200 ... 299
execution errors
300 ... 399
system errors
400 ... 499
query errors
1000 ... 9999
module dependent errors