![THORLABS PM5020 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/thorlabs/pm5020/pm5020_user-manual_1110985055.webp)
Page 51
PM5020 Dual-Channel Benchtop Power and Energy Meter
Chapter 4 Operating Instructions
Rev: 1.0, 18-Aug-2022
© 2022 Thorlabs
Chapter 5 Specifications
All are valid at 23 ± 5°C and 45 ± 15% rel. humidity (non condensing)
Item #
PM5020
Sensor Compatibility
Compatible Sensor Types
Photodiode Sensors S100C Series
Thermal Sensors S300C/S400C Series
Thermal Position Sensors S400C Series
Pyroelectric Sensors ES100C/ES200C/ES300C/ES400C Series
Photodiodes (Max 10 mA)
Thermopiles incl 4Q (Max 1 V)
Pyroelectric Sensors (Max 100 V)
Sensor Connector
DE-9 Female, Rear
Photodiode Sensor Current Input (CH1 and CH2)
Units
W, dBm, W/cm², lx, A
Measurement Range
21 Ranges:
2 nA to 10 mA
Ranges Selectable in W or A,
Sensor Dependent
Display Resolution
0.1 pA / Responsivity Value (A/W)
Measurement Uncertainty
±0.2% Full Scale (1 µA to 10 mA Ranges)
±0.5% Full Scale (2 nA to 400 nA Ranges)
Analog Bandwidth
DC - 250 kHz,
Dependent on Sensor and Settings
Wavelength Correction
nm [A/W]
Sensor Dependent
Thermopile Sensor Voltage Input (CH1 and CH2)
Units
W, dBm, W/cm², V
Measurement Range
9 Ranges:
2 mV to 1 V
Ranges Selectable in W or V,
Sensor Dependent
Display Resolution
0.1 µV / Responsivity Value (V/W)
Measurement Uncertainty
±0.5% Full Scale
Analog Bandwidth
DC - 1 kHz
Dependent on Sensor and Settings
Wavelength Correction
nm [V/W]
Sensor Dependent
Pyro Sensor Voltage Input (CH1 and CH2)
Units
J, J/cm², W, W/cm²
Measurement Range
12 Ranges:
20 mV to 100 V
Ranges Selectable in J or V,
Sensor Dependent
Display Resolution
1 µV / Responsitivity Value (V/J)
Measurement Uncertainty
±0.5% Full Scale
Max Repetition Rate
80 kHz
Wavelength Correction
nm [V/J]
Содержание PM5020
Страница 1: ...PM5020 Dual Channel Optical Power and Energy Meter 2022 User Guide...
Страница 2: ...Version Date 1 0 18 Aug 2022 Copyright 2022 Thorlabs...
Страница 65: ......