Page 48
MTN028837-D02
PM5020 Dual-Channel Benchtop Power and Energy Meter
Chapter 4 Operating Instructions
4.11.3.2.2 RS232 Operation
The power meter PM5020 can be operated with serial communication via RS232.
Connect the PM5020 power meter using an appropriate connector cable.
Go to the
and open the submenu Interface
.
The below screen "Serial" provides the information to connect the PM5020 to the network.
Please use a RS232 standard protocol for RS232 communications. For questions, please contact our engin-
eers at
Please set up the port as follows:
Parameter
Setting
Baud Rate
115.200 Bit/s (default)
Supported range of Baud Rate: 9.600 and 230400 Bit/s
Data Bits
8
Parity
None
Stop Bits
1
Flow Control
None
Termination Character
LF (x0A; \n). The termination character needs to be enabled.
Access to Dirver the SCPI condes to establish remote control software can be found int eh chapter Write
Your Own Application (
).