Troubleshooting
________________________________________________________________________________
______________________________________________________________________________________________________
© 2012 Thorlabs
Is the bias voltage set to 0 in photo-element mode?
Enter an appropriate bias voltage with the command
":VBIAS:SET <NR3>",
or in the channel menu
.
Is the correct photo diode efficiency
set (A/W)?
Enter the coefficient in the channel menu or with the command:
":CALPD:SET <NR3>"
Do you have chosen an appropriate measurement range for your application?
Use the command
":RANGE <NR1>"
or use the channel menu to
select a higher or lower sensitivity
If this does not help to resolve your trouble, please contact Thorlabs’ Tech Support.
(refer to section 5.6, “