LSK-GG
Chapter 5: Maintaining the LSK-GG
Rev A, November 11, 2016
Page 7
Chapter 5 Maintaining the LSK-GG
There are no serviceable parts in LSK-GG. If you suspect a problem with the LSK-GG, please contact our nearest
office for assistance from an applications engineer (see
Chapter 12 Thorlabs Worldwide Contacts
for details).
5.1. Cleaning
WARNING
Disconnect the power supply before cleaning the unit to prevent any injuries.
CAUTION
DO NOT expose the unit to spray, liquids, or solvent. This may damage the unit. DO NOT allow
spilled fluid to enter the internal mechanism. If spillage does occur, clean up immediately using an
absorbent tissue.
To clean the Controller Box and Scan Head:
Use a damp lint-free cloth to wipe the outer surface of the unit.
Use Compressed Air to remove the dust particles on the mirrors.
CAUTION
DO NOT touch the surface of the Galvo mirrors with fingers or tools to clean the mirrors.
5.2. Troubleshooting
Problem
Solution
X FAULT and/or Y FAULT LED turns on
Check all cable connections.
Make sure the Scan Head is free from any obstruction.
Unplug and replug the power supply.
Scanner has exceeded the maximum recommended
mechanical/electrical range. There are internal
mechanical stops within the scanner that prevent it from
spinning a full 360
, however during a fault state, the
current must be shut off before the rotor reaches these
internal stops.
Please contact Thorlabs Technical Support for assistance.
CAUTION
DO NOT let the system stay in this over-position
indefinitely. This may damge the scanner.
!
!
!
!
!
!
!
!
Содержание LSK-GG
Страница 1: ...LSK GG Laser Scanning Kit Galvo Galvo Scan Head User Guide...
Страница 22: ...www thorlabs com...