Kurios
®
Liquid Crystal Tunable Filters
Chapter 3: Description
Rev G, September 3, 2019
Page 5
3.2. Controls and Connections
3.2.1. KURIOS Controller: Front Panel
Figure 9
KURIOS Front Panel
Controls
WAVELENGTH:
Adjust the passband wavelength in MANUAL mode.
BANDWIDTH:
KURIOS-WB1(/M): Toggle between WIDE and BLACK (beam blocking) modes.
KURIOS-WL1(/M): Toggle between WIDE and BLACK (beam blocking) modes.
KURIOS-XL1(/M): Toggle between NARROW and BLACK (beam blocking) modes.
KURIOS-XE2(/M): Toggle between NARROW and BLACK (beam blocking) modes.
KURIOS-VB1(/M): Toggle between WIDE, MEDIUM, NARROW and BLACK (beam blocking) modes.
Connectors
TRIGGER IN:
Connect a 5 V TTL signal. It is used in the “SEQUENCE (with external trigger)” and “ANALOG (with
external trigger)” modes. In these modes, the TRIGGER IN is used to trigger the controller to switch
wavelength.
ANALOG IN:
Connect a 0 - 5 V analog signal. It is used in the ANALOG mode. In this mode, the filter adjusts its
wavelength, in accordance to the voltage of the analog signal. The ANALOG IN signal tunes the filter’s
passband wavelength from the shortest wavelength (0 V,
𝑊𝑠
) to the longest wavelength (5 V,
𝑊𝑙
),
meaning that the voltage step size for analog tuning is
V/nm.