![THORLABS KURIOS-VB1 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/thorlabs/kurios-vb1/kurios-vb1_user-manual_1110869019.webp)
Kurios
®
Liquid Crystal Tunable Filters
Chapter 5: Operation
Rev G, September 3, 2019
Page 15
5.2.3. Operation / Control Mode
Control Mode
Figure 17 Software Control Mode
Manual:
Set filter to manual mode. Tuning the knob at controller front panel and drag wavelength bar and
wavelength setting frame are available.
Sequence:
Set filter to sequence mode. There are two trigger modes available: Internal (default), External.
Internal
:
The sequence table is available and filter output wavelength switches according to the
sequence interval time in the sequence table.
External
:
The sequence table is available and filter output wavelength switches according to the
External trigger time.
Analog:
Set filter to analog mode. There are two trigger modes available: Internal (default), External.
Internal:
User inputs a 0-5V voltage from Analog in port. The switching time is according to the internal
time.
External:
User inputs a 0-5V voltage from Analog in port, filter output wavelength switches according
to the External trigger in time.
Bandwidth Mode
Figure 18 Software Bandwidth Mode
Different version of KURIOS tunable filter has different bandwidth mode available.
KURIOS-WB1(/M),KURIOS-WL1(/M) : WIDE, BLACK
KURIOS-VB1(/M): WIDE, MEDIUM, NARROW, BLACK
KURIOS-XE2(/M), KURIOS-XL1(/M): NARROW, BLACK
Wavelength
Figure 19 Software Wavelength
The wavelength bar indicates the current wavelength setting. Dragging the bar or inputting the wavelength in the
frame sets the working wavelength.