K-Cube
™ Liquid Crystal Controller
Chapter 3: Getting Started
Page 6
CTN014867-D02
3.2.3. Using the Baseplate
The baseplate must be bolted to the worksurface before the K-Cube is fitted, as shown below. The K-cube
is then placed on two dowels in the baseplate and secured by two clips.
Figure 4
Using The Baseplate
3.3.
Electrical Installation
3.3.1. Rear Panel
Figure 5
Rear Panel Connections
MOD IN
(SMA Female Connector) - external analog signal input, 0~+5 V DC;
AUX
(8 Position Circular Connector) - Reserved for future use
OUTPUT
(BNC Female Connector) - controller voltage output to driver Liquid Crystal Retarders, 0~+/-25 V
RMS.
3.3.2. Supply Voltage and Current Requirements
Supply
Minimum
Maximum
Max Operating
Current
+5V
+4.9V
+5.1V
350 mA
+15V
+14.5V
+15.5V
220 mA
-15V
-14.5V
-15.5V
50 mA